Computation of the electron beam quality kQ,Q0 factors for the NE2571, NE2571A and NE2581A thimble ionization chambers using PENELOPE

dc.contributor.authorErazo Caluqui, Fabián Gonzalo
dc.date.accessioned2018-01-11T16:47:56Z
dc.date.available2018-01-11T16:47:56Z
dc.date.issued2017-06-01
dc.description.abstractThe quality correction factor kQ,Q0 for electron beams was calculated for three thimble ionization chambers, namely, NE2571, NE2571A and NE2581A. The Monte Carlo code PENELOPE was used to estimate the overall correction factor fc,Q of these chambers for electron beams with nominal energies ranging between 6 and 22 MeV, corresponding to a Varian Clinac 2100 C/D. A 60Co beam was used as reference quality Q0. Also eight monoenergetic electron beams reproducing the quality index R50 of the Clinac beams were considered. The kQ,Q0 factors were calculated as the ratio between fc,Q and fc,Q0 . Those obtained for the NE2571 ionization chamber show a nice agreement with those calculated by Muir and Rogers with EGSnrc. As it occurred to other ionization chambers analyzed in previous works, the kQ,Q0 factors found for the monoenergetic beams are larger (smaller) than those corresponding to the Clinac beams at low (high) R50 values, the differences being slightly above 0.5%. Finally, the kQ,Q0 factors obtained in the case of the NE2571A chamber are systematically ?0.5% below those of its predecessor chamber, the NE2571.
dc.identifier.doi10.1016/j.ejmp.2017.05.053
dc.identifier.issn11201797
dc.identifier.urihttps://www.scopus.com/inward/record.uri?eid=2-s2.0-85020928193&doi=10.1016%2fj.ejmp.2017.05.053&partnerID=40&md5=2c15df5cc4094801710600ecd8dea4c3
dc.identifier.urihttp://dspace.ucuenca.edu.ec/handle/123456789/29280
dc.language.isoen_US
dc.publisherASSOCIAZIONE ITALIANA DI FISICA MEDICA
dc.sourcePhysica Medica
dc.subjectElectron Beams
dc.subjectKqq0 Correction Factors
dc.subjectNe2571
dc.subjectNe2571a
dc.subjectNe2581a
dc.subjectPenelope
dc.titleComputation of the electron beam quality kQ,Q0 factors for the NE2571, NE2571A and NE2581A thimble ionization chambers using PENELOPE
dc.typeArticle
dc.ucuenca.afiliacionerazo, f., instituto del cáncer – solca, cuenca, ecuador, escuela de tecnología médica, facultad de medicina, universidad de cuenca, cuenca, ecuador
dc.ucuenca.correspondenciaLallena, A.M.; Departamento de Física Atómica, Molecular y Nuclear, Universidad de GranadaSpain; email: lallena@ugr.es
dc.ucuenca.cuartilQ2
dc.ucuenca.embargoend2022-01-01 0:00
dc.ucuenca.factorimpacto0.679
dc.ucuenca.idautor1001784030
dc.ucuenca.indicebibliograficoSCOPUS
dc.ucuenca.volumen38

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
documento.pdf
Size:
168.92 KB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
19.94 KB
Format:
Plain Text
Description:

Collections