Logo Repositorio Institucional

Please use this identifier to cite or link to this item: http://dspace.ucuenca.edu.ec/handle/123456789/29280
Title: Computation of the electron beam quality kQ,Q0 factors for the NE2571, NE2571A and NE2581A thimble ionization chambers using PENELOPE
Authors: Erazo Caluqui, Fabián Gonzalo
metadata.dc.ucuenca.correspondencia: Lallena, A.M.; Departamento de Física Atómica, Molecular y Nuclear, Universidad de GranadaSpain; email: lallena@ugr.es
Keywords: Electron Beams
Kqq0 Correction Factors
Ne2571
Ne2571a
Ne2581a
Penelope
Issue Date: 1-Jun-2017
metadata.dc.ucuenca.embargoend: 1-Jan-2022
metadata.dc.ucuenca.volumen: 38
metadata.dc.source: Physica Medica
metadata.dc.identifier.doi: 10.1016/j.ejmp.2017.05.053
Publisher: ASSOCIAZIONE ITALIANA DI FISICA MEDICA
metadata.dc.type: Article
Abstract: 
The quality correction factor kQ,Q0 for electron beams was calculated for three thimble ionization chambers, namely, NE2571, NE2571A and NE2581A. The Monte Carlo code PENELOPE was used to estimate the overall correction factor fc,Q of these chambers for electron beams with nominal energies ranging between 6 and 22 MeV, corresponding to a Varian Clinac 2100 C/D. A 60Co beam was used as reference quality Q0. Also eight monoenergetic electron beams reproducing the quality index R50 of the Clinac beams were considered. The kQ,Q0 factors were calculated as the ratio between fc,Q and fc,Q0 . Those obtained for the NE2571 ionization chamber show a nice agreement with those calculated by Muir and Rogers with EGSnrc. As it occurred to other ionization chambers analyzed in previous works, the kQ,Q0 factors found for the monoenergetic beams are larger (smaller) than those corresponding to the Clinac beams at low (high) R50 values, the differences being slightly above 0.5%. Finally, the kQ,Q0 factors obtained in the case of the NE2571A chamber are systematically ?0.5% below those of its predecessor chamber, the NE2571.
URI: https://www.scopus.com/inward/record.uri?eid=2-s2.0-85020928193&doi=10.1016%2fj.ejmp.2017.05.053&partnerID=40&md5=2c15df5cc4094801710600ecd8dea4c3
http://dspace.ucuenca.edu.ec/handle/123456789/29280
ISSN: 11201797
Appears in Collections:Artículos

Files in This Item:
File Description SizeFormat 
documento.pdf168.92 kBAdobe PDFThumbnail
View/Open


This item is protected by original copyright



Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

 

Centro de Documentacion Regional "Juan Bautista Vázquez"

Biblioteca Campus Central Biblioteca Campus Salud Biblioteca Campus Yanuncay
Av. 12 de Abril y Calle Agustín Cueva, Telf: 4051000 Ext. 1311, 1312, 1313, 1314. Horario de atención: Lunes-Viernes: 07H00-21H00. Sábados: 08H00-12H00 Av. El Paraíso 3-52, detrás del Hospital Regional "Vicente Corral Moscoso", Telf: 4051000 Ext. 3144. Horario de atención: Lunes-Viernes: 07H00-19H00 Av. 12 de Octubre y Diego de Tapia, antiguo Colegio Orientalista, Telf: 4051000 Ext. 3535 2810706 Ext. 116. Horario de atención: Lunes-Viernes: 07H30-19H00