Title: Electron beam quality kQ,Q0 factors for various ionization chambers: a Monte Carlo investigation with penelope
Other Titles: Physics in Medicine and Biology
Authors: Erazo, F.
Brualla, L.
Lallena, A. M.
Keywords: Ionization
Electron
Issue Date: 2014
metadata.dc.ucuenca.paginacion: Páginas 6673-6691
metadata.dc.description.numberSequence: 
volumen 59: número 21
metadata.dc.identifier.doi: doi: 10.1088/0022-3727/59/21/6673
metadata.dc.description.city: 
Cuenca
metadata.dc.type: Article
Abstract: 
n this work we calculate the beam quality correction factor ${{k}_{ ext{Q},{{ ext{Q}}_{0}}}}$ for various plane-parallel ionization chambers. A set of Monte Carlo calculations using the code penelope/penEasy have been carried out to calculate the overall correction factor fc,Q for eight electron beams corresponding to a Varian Clinac 2100 C/D, with nominal energies ranging between 6 MeV and 22 MeV, for a 60Co beam, that has been used as the reference quality Q0 and also for eight monoenergetic electron beams reproducing the quality index R50 of the Clinac beams. Two field sizes, 10 × 10 cm2 and 20 × 20 cm2 have been considered. The ${{k}_{ ext{Q},{{ ext{Q}}_{0}}}}$ factors have been calculated as the ratio between fc,Q and ${{f}_{ ext{c},{{ ext{Q}}_{0}}}}$ . Values for the Exradin A10, A11, A11TW, P11, P11TW, T11 and T11TW ionization chambers, manufactured by Standard Imaging, as well as for the NACP-02 have been obtained. The results found with the Clinac beams for the two field sizes analyzed show differences below 0.6%, even in the case of the higher energy electron beams. The ${{k}_{ ext{Q},{{ ext{Q}}_{0}}}}$ values obtained with the Clinac beams are 1% larger than those found with the monoenergetic beams for the higher energies, above 12 MeV. This difference can be ascribed to secondary photons produced in the linac head and the air path towards the phantom. Contrary to what was quoted in a previous work (Sempau et al 2004 Phys. Med. Biol. 49 4427–44), the beam quality correction factors obtained with the complete Clinac geometries and with the monoenergetic beams differ significantly for energies above 12 MeV. Material differences existing between chambers that have the same geometry produce non-negligible modifications in the value of these correction factors.
URI: http://dspace.ucuenca.edu.ec/handle/123456789/22005
Appears in Collections:Artículos

Files in This Item:
File Description SizeFormat 
scopus 61.pdfdirección17.31 kBAdobe PDFThumbnail
View/Open


This item is protected by original copyright



This item is licensed under a Creative Commons License Creative Commons

 

Centro de Documentacion Regional "Juan Bautista Vázquez"

Biblioteca Campus Central Biblioteca Campus Salud Biblioteca Campus Yanuncay
Av. 12 de Abril y Calle Agustín Cueva, Telf: 4051000 Ext. 1311, 1312, 1313, 1314. Horario de atención: Lunes-Viernes: 07H00-21H00. Sábados: 08H00-12H00 Av. El Paraíso 3-52, detrás del Hospital Regional "Vicente Corral Moscoso", Telf: 4051000 Ext. 3144. Horario de atención: Lunes-Viernes: 07H00-19H00 Av. 12 de Octubre y Diego de Tapia, antiguo Colegio Orientalista, Telf: 4051000 Ext. 3535 2810706 Ext. 116. Horario de atención: Lunes-Viernes: 07H30-19H00